Nikon Live Cell 2 Manual - Página 4
Navegue en línea o descargue pdf Manual para Microscopio Nikon Live Cell 2. Nikon Live Cell 2 14 páginas.
XY speed
4-way display selector
(access to x,y,z
information)
Fig. 9
Incubation Chamber Control
Temperature and CO
control is facilitated through the external Oko-lab Touch controller (Fig. 11) which in turn allows
2
for monitoring of the incubator status.
1. The home screen shows a readout of the current temperature and CO
status indicator shows green when the settings are within the normal limits but will change to yellow if the
temperature/CO
accompanied by an audible alarm which can be silenced by pressing the alarm indicator on the touch screen.
2. Selecting temperature or CO
can be changed using the +/- buttons. Confirm the settings by pressing set.
3. For CO
, the valve on the tank needs to be adjusted to ensure that it is entering the system correctly: open the
2
top valve on the CO
̴ 0 .3 bar.
PFS on/off
z-speed
concentration drops too low for a significant amount of time. This yellow indicator is
2
opens a second window where the set point can be adjusted (Fig. 13) The values
2
bottle (anticlockwise), then turn the regulator knob (clockwise) to achieve a pressure of
2
Nikon Imaging Centre @ King's College London
Nikon
Nikon
Imaging Centre
Imaging Centre
LED lamp
switch
concentration (Fig. 12), the coloured
2
Fig. 10
Fig. 1
3
King's College London
King's College London
Lamp
intensity
controls
Fig. 10