dispomed MODUFlex Manual - Página 13

Navegue en línea o descargue pdf Manual para Equipamiento médico dispomed MODUFlex. dispomed MODUFlex 18 páginas. Veterinary anesthesia machine
También para dispomed MODUFlex: Manual de instalación (4 páginas)

dispomed MODUFlex Manual
DAILY LEAK
CHECK PROCEDURE
The unit's integrity should be verified every day, and at each
replacement of patient circuit components such as breathing
bags, tubes etc . A properly maintained anesthesia machine
reduces gas consumption and operating room pollution .
1
Connect the circuit to be verified to the fresh gas
outlet of the anesthesia machine .
3
Open the flowmeter to approximately 200 ml .
13
Prerequisites:
Oxygen supply of 40 to 60
psi (3 bar) Breathing circuit
and bag .
2
Obstruct the patient's circuit outlet with a plug or
your thumb .
4
Close the gas evacuation valve .
805-0975-030