DAILY LEAK
CHECK PROCEDURE
The unit's integrity should be verified every day, and at each
replacement of patient circuit components such as breathing
bags, tubes etc . A properly maintained anesthesia machine
reduces gas consumption and operating room pollution .
1
Connect the circuit to be verified to the fresh gas
outlet of the anesthesia machine .
3
Open the flowmeter to approximately 200 ml .
13
Prerequisites:
Oxygen supply of 40 to 60
psi (3 bar) Breathing circuit
and bag .
2
Obstruct the patient's circuit outlet with a plug or
your thumb .
4
Close the gas evacuation valve .