1-CUBE OXI Manual del usuario - Página 4

Navegue en línea o descargue pdf Manual del usuario para Instrumentos de medida 1-CUBE OXI. 1-CUBE OXI 11 páginas. Oxygen measuring device

5.1.2 CLEANING OF THE DEVICE:
Connect the device by fixing of inlet hose with the connection (4) on oximeter (see pic.2) to the water cock. In case
it is not possible to connect the inlet hose to the water cock please use the corresponding reduction – attach the
corresponding hose to the reduction. The inlet hose on the water cock (3) must be secured not to fall out during
rinsing. Place the outlet hose into the drain and open the control valve (2). Wash water goes through inlet hose to
the measuring chamber (3) and then it goes out of the chamber through the outlet hose to the drain.
Once the
cleaning is over switch off the water cock. Disconnect the device from the water cock and let the water go out from
the measuring chamber.
Clean the external surface with soft flannel moistened with water. After this handling the device is ready for next
measurement.
5.2 CONTROL OF THE ELECTRONICS
The device (i.e. switch-on, shutdown, measurement, saving of the measured sample in the memory) is controlled
with the help of the membrane keyboard and the device display.
Pic. 1 Keyboard with display
key
key
key
key
key
key ENTER
Display
mínus -
plus +
rotation
OFF
ON
5.2.1 SWITCH-ON AND SHUTDOWN OF THE DEVICE
Press the key ON– to switch on the device. Press the key OFF – to switch off the device (hold the key untill the the
device is switched off)
The device is automatically switched of after particular time of standstill of device – the text: „ENT.mer."
is displayed .