akrometrix DIC Manual del usuario - Página 12
Navegue en línea o descargue pdf Manual del usuario para Unidad de control akrometrix DIC. akrometrix DIC 19 páginas.
3. Press the Adjust Height button and lower the TherMoiré sample support so that
the level of the calibration target is the same as sample to be measured.
Typically this will be such that the bottom of the calibration target is within 5 mm
of the oven heater bulbs.
4. Adjust the right camera so that it is roughly centered on the calibration target
dot field.
5. Move the calibration target left or right and up or down to make its dot field fit
within the right camera's FOV.
6. Align the left camera such that it matches the right camera's FOV. There is a
hole in the Common Crossbar Mount support behind where the camera
attaches for the largest FOV. This allows for adjusting the left camera's angle
while the camera mount plate is securely attached using the thumbscrews.
Figure 2.6 Two Images of Calibration Target in DIC Surface Measurement
7. Adjust aperture and focus so that the cal target is adequately bright and in
focus.
8. To get best focus, open the iris fully on one camera lens and reduce the light
source intensity to compensate.
Page 12 of 19
Note: Only the FOV that can be seen in both cameras
will be useful for analysis. When connecting the
cameras to their support bracket, it is important to line
them up so that they image the same FOV.
40-5000-01500 Rev. A
DIC User Manual