amc VIBRO AVM 1002D Manuel de l'utilisateur - Page 12

Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} amc VIBRO AVM 1002D. amc VIBRO AVM 1002D 14 pages. Vibration monitoring system with relay outputs

amc VIBRO AVM 1002D Manuel de l'utilisateur
Picture 8-1 // Functions of configuration switches
Example:
Monitoring of the RMS of the vibration signal velocity, using 3 Hz high pass filter and 10 KHz
low pass filter, for 100 mm/s range and internal power loop.
The following configuration of the switches must be set:
SW1
SW2
OFF
ON
OFF
OFF
OFF
ON
OFF
ON
OFF
OFF
OFF
ON
ON
OFF
1
8
1
6
WARNING!
The set-up of the switches configuration should be done on a switched off device. If the set-up
was done on an operating module, it needs to be restarted in order to activate the new
configuration.
12
AV MONITOR 1002D