III. Turning the Instrument On
a.
Press power
button for three seconds: red alarm light/audible beep will come on briefly followed by a warm up cycle as follows:
Figure 3:
initial screen
b.
The instrument will start a warm-up cycle with a count down from "60" to "0.00". During this phase, the circulation pump will start and
continue to work until the
during detector operation.
c.
At the end of warm-up, the unit will emit an audible beep and the alarm light will flash briefly, then the screen will start displaying a
rotating fan symbol (Figure 5); this "ready screen" indicates the instrument is now ready to measure ozone gas levels.
d.
Use the Ozilla Ozone Gas Detector to detect ozone gas levels in rooms or cabinets where the Ozilla Ozone Sterilizer device is
operating, or in general to check ozone presence in desired areas. To detect gas levels, point the device intake nozzle (Figure 1)
towards the area where ozone gas is suspected. It is recommended that users try the Ozilla Ozone Gas Detector in ozone-free areas
to get background readings and possibly to zero the instrument if needed (Section III.1)
NOTE: to avoid false positive ozone readings, avoid areas or surfaces containing relatively high alcohol or chemical vapor content.
Brief background readings of up to 0.2 ppm are normal and should not cause alarm. Any sustained (10 minutes or more) readings of
over 0.5 ppm ozone gas should be investigated and remedied.
IV. Menu Settings and Adjustments
A. MAIN MENU MODE: To enter the menu mode, press the [OK] button for 3 seconds. The following screen appears
MAIN MENU
•
Basic Set
•
History data
•
Data Analyse
•
Zero cali
•
Range cali
•
Alarm param
•
DETC&STOR SET
•
Date Time
•
Reset factory
B. BASIC SETUP: To enter this mode, press the down button to highlight the "Basic Set" option, then press [OK] button. The following
screen is displayed:
Basic Set
• Range&Unit
• Channel Set
• Address Set
• Bump Flow
• Tempr cor
• Humid cor
• Language
English
To change a particular option, navigate with the up
detector proper functioning, It is neither needed nor recommended to make any changes to the Basic Setup options. Available
Ozilla® Ozone Gas Detector Pro
.
Figure 4: Detector warms up by
counting down from 60 to 0 seconds
button is pressed, or until the detector is turned off. Do not block either intake or exhaust ports
RUN
STOP
Main menu visible screen is highlighted with a
thick border. Use the down
selections below the highlighted box,
Basic Setup initial screen view in highlighted
with a thick border. Use the down
255
view selections below the highlighted box,
35
Figure 5:
instrument is ready to detect ozone gas
button to view
button to
or down
buttons to the desired item, then press the [OK] button. For
"ready screen" indicates