amsbio Ozilla AMS.E4002GD Manuel de démarrage rapide - Page 5

Parcourez en ligne ou téléchargez le pdf Manuel de démarrage rapide pour {nom_de_la_catégorie} amsbio Ozilla AMS.E4002GD. amsbio Ozilla AMS.E4002GD 6 pages.

language options are English and Chinese.
C. ALARM PARAMETERS SETUP: To enter this mode from the MAIN MENU screen, press the down button to highlight the "Alarm
param" option, then press the [OK] button. The following screens are displayed:
Alarm Param
O3
Primary "Alarm Param" screen
After pressing the "Alarm Param" option in MAIN MENU, a primary "Alarm Param" screen appears, where "SET" is automatically
highlighted. Press the [OK] button to access the secondary "Alarm Param" screen.
There are two alarm parameter settings, one for a low alarm threshold and one for a high alarm threshold.
To set a low alarm threshold value, first set the "Alarm 1" value by clicking the [OK] button. Use the up
change the PPM value, then press [OK] again. Next, use the down button to highlight the Mode selection then click [OK]. The HI ALR
(default) in the window will blink. Use the up or down buttons to change to "LO ALR" then click [OK]. The detector will give an audible
alarm when the ozone gas reading is below the low alarm value chosen.
To set a high alarm value, navigate screen with up
PPM to the value desired (up to 50.00 PPM) then press the [OK] button again. The detector will now give an audible alarm when the
ozone gas reading is higher than the high alarm value chosen.
D. DATE AND TIME SETUP: To enter this mode from the MAIN MENU screen, press the down button to highlight the "Date Time" option,
then press the [OK] button. The "Date Time" screen is displayed as follows:
MAIN MENU
Basic Set
History data
Data Analyse
Zero cali
Range cali
Alarm param
DETC&STOR SET
Date Time
Reset factory
Use the up
or down
Use the up
or down
When done, press the [Back] button to return to the MAIN MENU screen, then press [Back] again to go to the "ready screen" if
needed.
Ozilla® Ozone Gas Detector Pro
Alarm Param
Sensor
Alarm 1:
SET
Mode:
Alarm 2: 10.00
Mode:
Secondary "Alarm Param" screen
or down
Date Time
2017-01-01
Date
Time
12:00:00
buttons to navigate to highlight the digit you want to change, then press [OK].
buttons to change the digit to the desired value, then press OK.
O3
5.00
PPM
HI ALR
PPM
HI ALR
buttons to the Alarm 2 option then press the [OK] button. Change the
or down
buttons to