Alemlube O30060 Manuel d'instruction - Page 10
Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} Alemlube O30060. Alemlube O30060 17 pages. Oil pump
Procedure for entering the correction factor directly
Carefully follow the procedure indicated below.
Formula: Proper correction factor = current correction factor × (actual value / display value)
Example: Actual value
Display value
Current correction factor
Proper correction factor
1
Wait for the meter to go standby.
2
Reset the resettable total.
3
Press the MOVE key. Keep it pressed until it shows similar
to the right image (the digit flash in
the meter is under the calibration mode.
4
Press the RESET key to choose the right digit from 0 to 9.
Press the MOVE key to start the next digit. So the digit of
correction factor can be changed one by one.
5
Make sure the correction factor is right, press the MOVE key.
Keep it pressed until the indication of calibration mode quits
and the factor is saved.
NSW/ACT
QLD/PNG
TEL: (02) 9939 0711
TEL: (07) 3889 8480
FAX: (02) 9939 0411
FAX: (07) 3889 8490
OWNER'S TECHNICAL MANUAL
Oil Transfer Kit with Meter & Trolley
20.75
18.96
1.000
1.000 × (20.75 / 18.96) = 1.000 × 1.094 = 1.094
zone), which means
VIC/TAS
TEL: (03) 8787 8288
FAX: (03) 8787 8266
Visit our website at
www.alemlube.com.au
WA
SA/NT
TEL: (08) 9209 3066
TEL: (08) 8241 7111
FAX: (08) 9209 3933
FAX: (08) 8241 7011
or
www.alemlube.co.nz
O30060
NZ
TEL: (09) 447 1007
FAX: (09) 447 1008
10