Procedure for entering the correction factor directly
Carefully follow the procedure indicated below.
Formula: Proper correction factor = current correction factor × (actual value / display value)
Example: Actual value
Display value
Current correction factor
Proper correction factor
1
Wait for the meter to go standby.
2
Reset the resettable total.
3
Press the MOVE key. Keep it pressed until it shows similar
to the right image (the digit flash in
the meter is under the calibration mode.
4
Press the RESET key to choose the right digit from 0 to 9.
Press the MOVE key to start the next digit. So the digit of
correction factor can be changed one by one.
5
Make sure the correction factor is right, press the MOVE key.
Keep it pressed until the indication of calibration mode quits
and the factor is saved.
NSW/ACT
QLD/PNG
TEL: (02) 9939 0711
TEL: (07) 3889 8480
FAX: (02) 9939 0411
FAX: (07) 3889 8490
OWNER'S TECHNICAL MANUAL
Oil Transfer Kit with Meter & Trolley
20.75
18.96
1.000
1.000 × (20.75 / 18.96) = 1.000 × 1.094 = 1.094