Absolute Measurement CLT Manuel d'instruction - Page 6
Parcourez en ligne ou téléchargez le pdf Manuel d'instruction pour {nom_de_la_catégorie} Absolute Measurement CLT. Absolute Measurement CLT 9 pages.
Transducer Instruction Manual
Range:
Ultra
Type:
CLT
Revision:
V1.0
2.8
Sensing Roller Installation
The transducer has a choice of adaptors with bores of 17, 20, 25, 30 and 35mm diameter, to accept a
shaft to which a tension sensing roller or guide pulley can be mounted as shown below. Two set screws
at 90 degrees are used to hold the shaft in place. The tension sensing roller or guide pulley should be
dynamically balanced when the roller speed exceeds 300 RPM as excessive vibration can damage the
transducer. Any out of balance force should not exceed 10% of the maximum working force (MWF) of the
transducer.
The recommended maximum roller length is 450mm to avoid an excessive bending moment being
applied to the transducer and roller deflection.