Additional functions
File Menu:
User can save up to 10 full configurations of the
instruments and recall them whenever.
Configurations can be saved and recalled in
function of user and of use. Configurations include
all programming done on instrument, as the range.
Contrast adjustment:
Screen's contrast can be adjusted whenever to fit
with measurement environment
Screen Backlighting:
Time of backlighting can be programmed to save
battery
Autonomy:
Calys 50 autonomy is 8 hours in the worst condition
of use
Scaling:
In measurement and simulation, scaling allows
process signals to be displayed in % of FS or in all
other unit
This function also allows sensors to be corrected
after a calibration
Relative measurement:
•
Programming of a reference value different
from the one of the instrument (NUL function)
•
Substracting of constant value by measuring or
programming it from a measured value(TARE
function)
Square root:
In current measurement and simulation, this
function allows to take into account a quadratic
signal coming from transmitter of type ∆P
Statistical
functions:
Average, minimum, maximum and also number of
measurements done are always displayed.
Reset key allows values to be updated.
Calys 50 is supplied in standard with 6 testing leads, a quick battery charging system, an instruction manual
Instructions to order:
On-site calibrator
Protective carrying Case:
Ed 02/10/04
CALIBRATION
Simulation Menu:
Simulation value is set by entering value on keypad
or by changing the according digit with the cursor
Ramps generation:
Starting, ending and length time values of simple or
cyclic ramps can be set to do simulation.
Number of ramps can also be adjusted in case of
cyclic ramps for any signals.
Steps simulation:
2 modes are proposed.
•
•
In current simulation, user will have some additional
preset values in function of range and according to
0%, 25%, 50%, 75% and 100% from selected
gauge. Choice is done between gauges:
•
•
Synthesizer:
With 100 values manually set, Calys 50 allows
curve generation to be remade.
Transmitter function:
CALYS 50 is able to be used as a transmitter.
Measurement input is copied on the output with
scaling.
CALYS 50
ACL6050
AOIP
BP 182
91006 EVRY CEDEX
FRANCE
www.aoip.com
Program mode: Starting value, number of
steps and the length time have to be set
Manual mode: User has about a hundred of
preset values
0-20mA: linear or quadratic
4-20mA: linear or quadratic