AOIP CALYS 50 Calibrage - Page 4
Parcourez en ligne ou téléchargez le pdf Calibrage pour {nom_de_la_catégorie} AOIP CALYS 50. AOIP CALYS 50 6 pages. On-site multifunction calibrator
Functions and performances:@23°C ±5°C
Accuracies are given in % of reading(a CALYS 50 display) + fixed value
Direct current: Measurement
Calys 50 is able to measure up to 50 mA, with or
without power supply of loop (24V power supply).
Range
Resolution
Accuracy / 1yr
±50mA
1 µA
0,018%R + 2 µA
For measurements of transmitters outputs, special
ranges give a dual display using mA and % of full
scale.
Calys 50 also allows to linearize following linear or
quadratic signal.
In current measurement Hart compatibility can
selected to measure currents coming from Hart
protocol transmitters
Direct voltage: Measurement
Range
Resolution
Accuracy / 1yr
±100mV 1 µV
0,013%R + 3 µV
±1V
10 µV
0,013%R + 20 µV
±10V
100 µV
0,015%R + 200 µV
±50V
1 mV
0,015%R + 2 mV
Rin: input resistance
Frequency and counting: Measurement
Range
Resolution
20 kHz
< 0.01 Hz
Threshold triggering: 1V
Unit scale: pulse/min or Hz
Measurement on frequency signal and on dry
contacts
Measurement for counting will be done on defined
time or on infinite time.
Resistance: Measurement
Range
Resolution
Accuracy / 1yr
400 Ω
1 mΩ
0.012% R+ 10 mΩ
4000 Ω 10 mΩ
0.012% R+ 100 mΩ
2, 3 or 4 wires resistance measurement :automatic
recognition of number of connected wires, with
indication on screen
Ed 02/10/04
CALIBRATION
Remarks
Rin < 25 Ω
Remarks
Rin > 10 MΩ
Rin > 10 MΩ
Rin = 1MΩ
Rin = 1MΩ
Accuracy / 1yr
0.005%R
Remarks
Measureme
nt current =
0.25 mA
Measureme
nt current =
0.25 mA
DC current: Emission
Range
Resolution
24mA
1 µA
Emission with or without loop supply (24V)
Preprogrammed steps
0%
25%
4-20mA linear
4
0-20mA quad
0
4-20mA linear
4
0-20Ma quad
0
1,25
4-20mA valves
3,8-4 -4,2
Direct voltage: Emission
Range
Resolution
Accuracy / 1yr
100mV
1 µV
0,013% R+ 3 µV
2V
10 µV
0,013% R+ 30 µV
20V
100 µV
0,015% R+ 300 µV
Frequency and pulses: Emission
Range
Resolution
1000 Hz
< 0.01 Hz
10 kHz
10 Hz
Unit scale: pulse/min or Hz
Pulse emissions
Dry contact simulation
Max amplitude: 20V selectable by user
Resistance: Emission
Range
Resolution
Accuracy / 1yr
400 Ω
10 mΩ
0.014% + 30 mΩ
4000 Ω 100 mΩ
0.014%R + 300 mΩ
Resistance emission: establishing time <1ms for
compatibility with smart transmitters type
Accuracy / 1yr
0,018%R + 2 µA
50
75%
100%
%
8
12
16
20
5
10
15
20
5
8
13
20
5
11,25
20
12
19,20,21
Remarks
Load 1KΩ
Load 2KΩ
Load 4KΩ
Accuracy / 1yr
0.005% R
0.005% R
Remarks
Iext from
0.1 to 10
mA
Iext from
0.01 to 0.1
mA
Document susceptible d'être modifié