AOIP CALYS 50 Calibrage - Page 3
Parcourez en ligne ou téléchargez le pdf Calibrage pour {nom_de_la_catégorie} AOIP CALYS 50. AOIP CALYS 50 6 pages. On-site multifunction calibrator
"easy connect" system
This unique system is used by pushing on the
terminal's top, by inserting:
-
Wires with a diameter up to 3 mm,
-
Compensated thermocouple connectors,
-
Pin terminal on front panel,
and next by releasing.
Wires are tighten between 2 brass plates which
provide an great thermal gradient, so that allows a
very good cold junction compensation for
thermocouples.
Calys 50 allows 4mm connectors and also security
connectors to be connected on the front panel.
Display resolution:
Calys 50 allows the digit number after the dot to be selected: This function is justified by the needs of users
who want or not to display the best resolution for calibration or on the contrary limit it for simple verifications.
Functions
:
Calys 50 allows the following physical values to be measured and simulated:
•
Voltage
•
Current
•
Resistance
•
Temperature by resistive probes and thermocouples
•
Frequency for frequency signals and for dry contacts
It is able to perform scaling of process signals and so to correct temperature probes.
It is compatible with HART transmitters by inserting a 250ohms resistance which doesn't disturb digital data
transfer.
Display
Calys 50 dual display indicates permanently the
measurement value, and also the emitted value, the
gauge and the used functions.
On the top date, time and also external temperature
are also indicated.
During measuring average, maximum, minimum
and the number of measurements are displayed on
the left. While for emission this part of screen
displays all details of ramps, steps and constant
value emission functions.
Drop-down menus are used with the navigator, and
an on-line help is available to make easier
connections of probes and wires.
Ed 02/10/04