Hydraulic Pressure Test Pump User s Manual
Please download the latest version from
Configuration & Pressure drawing
1
2 3 4
5
8
7
Troubleshooting
Problem
Difficult to increase pressure with
pre-pressurization handle
Hard to fine adjust
it is difficult to turn valves or quick
connectors
O-Rings for pressure connector
P/N
1611300004
1611300220
1611300024
ADT927
'
[Version number:1409V12]
www.additel.com
1
-
Quick connector
2 - Reservoir release valve
(open it to connect with atmosphere)
3 - Upper cover of reservoir
4 - High-pressure pressurization and
6
fine pressure adjustment handle
(clockwise to increase pressure)
5 - Liquid reservoir
(inner is pressure media)
6 - Pre-pressurization handle
(clockwise to increase pressure)
7 - Screw of fitting cover
8 - Vent valve
(Open to connect with atmosphere)
Cause
A. Vent valve is not closed.
B. The O-ring seal is loose or damaged.
C. Not enough media is in the reservoir.
D. The pump is not horizontal.
A. The vent valve is not closed.
B. The gauges are not connected tight.
C. The
sealing ring
is aged or frayed.
D. The thread surface is not smooth
.
E. The connector type is mismatched to the gauge pressure port.
A. Too much force was previously applied.
B. The pre-pressurization handle may not adjust easily under high
pressure.
C. The threads have no lubrication
.
Size
4X1.5
6.5X3
6X2
Warnings and cautions
Operate the pump in the rated pressure range and do not exceed the
safety pressure of 15,000 psi (1,000 bar)
Close the valves and tighten the ports when transporting the pump.
Always open the reservoir release valve during operation.
Do not over tighten the connectors or handles to avoid any damage.
Change contaminated media immediately.
Do not let the media level get below the mid line of the reservoir when
the pump is place horizontal.
Keep the outer threads clean.
Additel is not liable for any safety problems or damages caused by
misuse or incorrect operation.
Specification
Pressure range:
12.5 psi ( 0.85 bar ) vacuum to
10,000 psi ( 700 bar ) positive pressure
Remark:
If local atmosphere pressure is 1 bar, the vacuum can reach to 0 85 bar;
If local atmosphere pressure is P, the vacuum can reach to
Temperature:
0~50)℃
Humidity:
< 95%RH
Adjustment resolution:
Safety pressure:
< 15,000 psi ( 1000 bar )
Pressure media:
Oil or deionized water
To reach the best performance oil is recommended
Recommended oil Di ethylhexyl Sebacate Isovoltine
Size:
Height: 5.31'' ( 135 mm );
Base: 11.42'' ( 290 mm ) x 7.80'' ( 198 mm )
Weight:
7.1 lb ( 3.2 kg ) ( excluding the media )
Pressure drawing
2
5
Close the vent valve.
Replace the O-ring
Fill with more fluid. Make sure fluid level is high enough after the
gas purging process.
Place the pump horizontally to prepare for pressure or vacuum.
Close the vent valve.
Tighten the gauges.
Replace sealing ring.
Use Teflon tape on the thread and turn it in tight.
Use the correct adapter.
Do not over tighten.
Adjust pressure by using fine adjustment handle. (Pressure may
need to be decreased with the fine adjustment handle before the
pre-pressure handle can be used).
Lubricate the thread.
M10X1, 1/8BSP, 1/8NPT
M20X1.5, 1/2BSP, 1/2NPT
M14X1.5, 1/4BSP, 1/4NPT, 3/8BSP
.
( P 85% )
0.015 psi ( 0.001 bar )
,
:
-
,