QNix
5500
®
7.2.2. Probe Electrical Data
The probes may only be used with the gauges QNix® 9500 and QNix® 5500.
7.3. Measuring Data
Measuring Data
Principle of Measurement
As per standard (varies by model)
Status indication
Probe type
Measuring range
Measuring interval
Units
Display resolution
Measurement uncertainty*
(based on factory standards)
Minimum measuring area (Ø)
Minimum curvature radius, convex
Minimum thickness of the base material
* Accuracy may deviate outside a range of -5 °C to 65 °C, but not more than the factor 2 of the
measurement uncertainty..
14
Fe
Magnetic field principal
Magnetic field change,
Hall effect
DIN EN ISO 2808,
DIN EN ISO 2178
DIN 50981, DIN 50984, BS 5411 (3 & 11), BS 3900-C5,
ASTM B 499, ASTM D 1400, ASTM D 1186, ASTM D 7091
by RGB-LED at the probe
See probe data sheet
Varies depending on probe type
≥ 100 measurements / min.
μm / mil (setting in battery compartment)
0,01 μm
± (1 μm + 2 %) to 2 mm
± 3 % from 2 mm
14,5 mm
6 mm
100 μm
NFe
Magnetic field principal
Eddy current
DIN EN ISO 2808,
DIN EN ISO 2360
± (2 μm + 2 %) to 2 mm
± 3 % from 2 mm
14,5 mm
50 mm
50 μm