iHealth View Manuel de l'utilisateur - Page 11

Parcourez en ligne ou téléchargez le pdf Manuel de l'utilisateur pour {nom_de_la_catégorie} iHealth View. iHealth View 12 pages. Wireless blood pressure wrist monitor

iHealth View Manuel de l'utilisateur
BP7S is intended for use in the electromagnetic environment specified below. The user of BP7S should ensure that it is used in such an
environment.
IMMUNITY test
Conducted RF
3 Vrms 150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m 80 MHz to 2.5
IEC 61000-4-3
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which BP7S is used exceeds the applicable RF compliance level above, BP7S should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating BP7S.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
portable and mobile RF communications equipment and the Wireless Blood Pressure Wrist Monitor
BP7S is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of BP7S can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and BP7S as recommended below, according to the maximum output power of the communication equipment.
Rated maximum output
power of transmitter
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
Complian
IEC 60601test level
ce level
3 V
3 V/m
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between
Table 3
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no
closer to any part of BP7S, including cables, than the recommended
separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
=
1
2 .
d
P
=
1
2 .
d
P
80 MHz to 800 MHz
=
2
3 .
d
P
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
Table 4
Separation distance according to frequency of transmitter
a
should be less than the compliance level in
b
m