Gambro Prisma System Instructions d'étalonnage - Page 2
Parcourez en ligne ou téléchargez le pdf Instructions d'étalonnage pour {nom_de_la_catégorie} Gambro Prisma System. Gambro Prisma System 2 pages. Scale diagnose and calibration procedures
place either or both of the 2600-gram weights on the scale and monitor the
values. The reading should be + 7 grams for either weight.
CALIBRATE SCALES
!
Access Service Mode per procedure above
!
Press the Calibrate softkey to access calibration screens and procedures for scales,
pressure sensors, setting filter clotting limits, setting the real time clock/calendar, and
setting the PRISMA ID number (for a complete description of these screens, consult the
Prisma Service Manual, pages 5-4 through 5-10).
!
Press Scales softkey to access calibration screen for scales. Note that the softkeys
appear on the screen in the approximate location of each scale on the machine (i.e., the
dialysate scale is on the left side, effluent scale is in the middle, and the replacement
scale is on the right side of the machine).
!
Press the softkey for the scale you wish to calibrate. Verify that no weight is applied to the
selected scale.
The first screen will instruct the user to "place weight on selected scale. Use
o
weight of –100 to 100 grams". DO NOT place any weight on the selected scale.
Wait for the "Scale Stable" message to appear.
Press Next softkey
o
The second screen will instruct the user to "place weight on selected scale. Use
o
weight of 2500 grams to 2700 grams". Place one weight (2600 grams) on the
selected scale. Wait for the "Scale Stable" message to appear.
NOTE: Place weight on middle hook of selected scale to insure even distribution
of weight
Press Next softkey
o
The third calibration screen will appear and instruct the user to "place weight on
o
selected scale. Use weight of 5100 grams to 5300 grams". Place both Prisma
weights on selected scale. Wait for "Scale Stable" message to appear.
NOTE: Place weights on outer hooks of selected scale to insure even distribution
of weight
Press Store softkey
o
!
Now that the selected scale is calibrated, the Service-Scales Calibrate screen returns to
the Service Scales screen. Remove the weights and calibrate any of the other scales
using this same procedure.
!
When scale calibrations are complete, press Exit softkey until the main Service screen
appears. Repeat DIAGNOSE SCALES to insure all scales are in calibration.
!
Press Exit softkey until main Service screen appears. Press Restart softkey to restart the
Prisma System. The machine will enter Setup Mode, Choose Patient screen.
CLINICAL ASSISTANCE SERVICE
!
Available 24 hours/day, including weekends and holidays
!
Call if further assistance is required
Reference:
Gambro Renal Products, Inc. Prisma System Service Manual, 1999, Chapter 5, pages 5-1
through 5-13.
10810 West Collins Avenue
Lakewood, CO 80215-4434
800-525-2623 or 303-232-6800
www.usa-gambro.com
2002 Gambro Renal Products, Inc. P/N 306100068 Rev A