360 YIELD CENTER 360 SOILSCAN Instructions d'installation - Page 8

Parcourez en ligne ou téléchargez le pdf Instructions d'installation pour {nom_de_la_catégorie} 360 YIELD CENTER 360 SOILSCAN. 360 YIELD CENTER 360 SOILSCAN 12 pages.

+
360 SOILSCAN
STORAGE GUIDELINES
Below are tips for storing your machine for more than 24
hours or over the winter.
+
Store the NO3-N Sensor (gray) dry. Use the
NO3-N Storage Bottle that came with the
sensor to cover the tip. Switch to this storage
method if the machine will not be used within
the next 24 hours.
Fill the small Reference Electrode Storage Bottle
+
¾ full with Standard Solution. Add cap to bottle
and insert over electrode (red), making sure
electrode tip is immersed in solution. Tighten
the cap on the bottle. Switch to this storage
method if the machine will not be used within
the next 24 hours.
+
Place the 360 SOILSCAN™ machine and sensors
in a mild environment (60°F to 80°F). Do not
leave the machine in an unheated shop as
exposure to cold temperatures can affect the
sensor performance. A good rule of thumb is to
treat the machine and sensors similar to how
you would treat your iPad.
+
Standard Solution can be stored at room
temperature.
+
Extra tips that have not been opened should
also be stored at room temperature in their
original packaging.
+
For winter storage, make sure the water tank on
your machine is empty and the water line
is purged.
SETUP INSTRUCTIONS
1
Figure 6: Reference Electrode (1) (Red): Always store wet in the
Standard Solution. NO3-N Sensor (2) (Gray): Store dry.
360YIELDCENTER.COM
360 SOILSCAN™
2