ASSEMBLY

}Each standard set can be assembled by simply attaching the AC adapter and power cord.
@
5
17
²
³
|
Fig. 16
6
7
Fig. 17
Connecting the AC Adapter and
1
Power Cord
# Be sure to use the provided AC adapter Use of other AC adapter
may cause a malfunction.
The power cord and AC adapter cord are vulnerable when bent or
twisted. Never subject it to excessive force.
Make sure that the main switch 1 is set to "
necting the power cord.
Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to the
section "PROPER SELECTION OF THE POWER SUPPLY CORD" at
the end of this instruction manual.
1. Connect the output connector 2 of the AC adapter to the input connec-
tor 3 on the rear of the microscope.
Do not tilt the microscope to the rear. Doing so could damage the
output connector 2.
2. Connect the power cord's connector 4 to connector 5 firmly.
Be sure to supply power from a grounded, 3-conductor power outlet
using the proper power cord. If the power outlet is not grounded
properly, Olympus can no longer warrant the electrical safety perfor-
mance of the equipment.
3. Connect the power cord's plug 6 to a wall power outlet 7. (Fig. 17)
(Figs. 16 & 17)