Using the Exciter Balancer U-EXBABG/EXBAUB/EXBAUG
}When observing fluorescence generated with U/B/G multiple excitation using double-dye fluorescence excitation and a
barrier filter, the exciter balancer can be used to adjust the balance between the intensities of the excitation light of the
fluorescent dyes.
Engagement adjustment lever
Pushed in: IN.
Pulled out: OUT.
Rotation: Adjustment (45°)
17
²
³
@
Fig. 18
Clamping screw
Attaching Procedure
1. Remove the dummy from the exciter balancer insertion slot.
2. Set the balancer's engagement adjustment lever @ in the vertical position
and insert the balancer into the insertion slot ² on the right side of the
illuminator.
3. Tighten the clamping screw ³ using the Allen screwdriver to fix the balancer.
Operating Procedure
1. Mount a commercially available double-dye fluorescence excitation filter
and a barrier filter in the slider position (see page 16 for the attaching
method).
2. Rotate the turret to engage these filters in the light path.
# Due to the characteristics of G-excitation, its brightness adjustment
range is lower than that of the B- or U-excitation. The brightness
adjustment range is also variable depending on the variance in
characteristics of the filter.
# Due to the filter rotation angle and the variance in characteristics of
the filter, irregularities in illumination may be observed on the top
and bottom of the visual field. But these irregularities do not affect
the area captured in photomicrography.