DigiFLO HH Manuel - Page 3
Parcourez en ligne ou téléchargez le pdf Manuel pour {nom_de_la_catégorie} DigiFLO HH. DigiFLO HH 5 pages. Concentrator analyzer
Pressure Unit Conversions
KPA
PSI
KPA
MMHg
KPA
CMH
O 10.2 CMH
2
Caution
"Caution: Do not expose this product to
pressures greater than 40 PSI, or damage
may result to internal components.
Do not use this product on pure oxygen from
high pressure gas tanks or liquid systems.
Use the DigiFLO Concentrator Analyzer
with unhumidified gas only. Use of this
product with water vapors can cause
erroneous readings and internal damage.
Indication For Use:
The DigiFLO Concentrator ANALYZER is a tool used by service personnel to measure Oxygen purity
and Flow Rate produced by an oxygen concentrator. Also, it measures gas pressure. Unless a patient is
specifically trained, it is not intended to be used by patients who are prescribed oxygen. The DigiFLO
Concentrator ANALYZER is intended to be used in an environment where oxygen concentrators are
being serviced or repaired. This includes Hospitals, Nursing Homes, Extended Care Facilities, Patient
Homes, and Respiratory Device Service and Repair Centers.
1 PSI = 6.9 KPA
7.5 MMHg = 1 KPA
O = 1 KPA
2
Instructions for use
1. Press and release the front button. Display will come on
in ●O2 Mode.
2. To reach one of the following modes, press and release
the button until LCD displays one of the following:
Mode
Units
●O2
%
●FLO
LPM
●KPA
KPA
3. Connect 1/8"ID x 1/4"OD silicone tubing to
ANALYZER barbed inlet.
In order to measure pressure, plug the outlet port with your
thumb.
3
Parameter
Oxygen Concentration
Flow Rate
Pressure
Gas Inlet
Mode LED
Outlet