DigiFLO HH Manuel - Page 5
Parcourez en ligne ou téléchargez le pdf Manuel pour {nom_de_la_catégorie} DigiFLO HH. DigiFLO HH 5 pages. Concentrator analyzer
1. Flush HH with Air.
2. Prepare a On / Off switchable 4.0 LPM, 41 KPA, 100% O2 flow source and Connect to
HH as follows:
1/8" ID Silicone Hose
100% O2
source
3. Power HH On and continue holding the button until HH displays CAL and then CAL is
replaced with O2 reading. Release the button.
4. Turn O2 Flow on.
5. Press the front button. Display will show FLO. Continue holding down the button until
HH displays CAL and then CAL is replaced with Flow Rate reading. Release the button.
6. Turn O2 Flow off.
7. Plug HH outlet and Turn Flow On. Thus 41 KPA pressure will apply to HH.
8. Press the front button. Display will show KPA. Continue holding down the button until
HH displays CAL and then CAL is replaced with Pressure reading. Release the button.
9. Unplug HH outlet.
10. Wait for ½ second.
11. Turn O2 Flow off.
12. Wait for ½ second.
13. Press the front button. Display will show O2. Continue holding down the button until HH
displays CAL and then CAL is replaced with O2 reading.
Warning
User should be cautious such that this product, does not perform inadvertent Field calibrations.
Cleaning
This product should be cleaned by applying a damp cloth or sponge to the outside case, such that
the display window is clean and transparent . Do not immerse the analyzer in any water solution
or any other cleaning/disinfecting agent.
Never clean the meter with organic solvents such as alcohol or acetone.
Field Calibration
5
DigiFLO confidential
information
1/8" ID Silicone Hose
Valve
HH