Siemens SIREMOBIL Iso-C Maintenance Instruction - Page 22
Parcourez en ligne ou téléchargez le pdf Maintenance Instruction pour {nom_de_la_catégorie} Siemens SIREMOBIL Iso-C. Siemens SIREMOBIL Iso-C 46 pages.
Également pour Siemens SIREMOBIL Iso-C : Instructions d'installation et de réglage (4 pages), Manuel d'instructions de service (16 pages), Manuel d'instructions d'installation et de réglage (6 pages), Manuel d'instructions d'installation (6 pages)

22
1.10
Technical Safety Checks (TSC)
Abbreviation: TSC = Technical Safety Checks
NOTE
Technical safety checks, test certificate
of the check
1 Identification data
2 General checks
2.1 Visual inspection: Is there any damage
affecting safety?
2.2 Are all cables and cable guides secure
and without any visible damage?
SIREMOBIL Iso-C
•
Within the purview of DIN VDE 0751-1, the operator of medical
engineering products has to perform technical safety checks at
regular intervals.
•
The checks listed in these maintenance instructions contain all
technical safety checks according to DIN VDE 0751-1 except for
the check of the completeness, presence and legibility of the
necessary operator documents.
•
A separate report with print number SP00.000.834.01... is avail-
able for each technical safety check.
•
Within the purview of DIN VDE 0751-1, we recommend docu-
menting the maintenance results both in the maintenance re-
port and in the TSC report. The reports should be filled out
completely and handed over to the client after maintenance is
complete.
•
The table below assigns each technical safety check listed in
document SP00.000.834.01... to the corresponding mainte-
nance work.
SPR2-230.101.02.06.02
07.05
General Information
Maintenance instructions for the
check/comments
Please fill in the fields.
n.a.
3. Mechanical safety, checks:
SIM Covers
SIM Cable deflectors
SIM Foot brake
SIM Brakes
SIM C-arm
Emergency Stop switch
SIM Monitors/displays
3. Electrical safety, check:
SIE Cables and plugs
SIE Voltage discharge rubber
Page 22 of 46
CS PS 24
Siemens AG
Medical Solutions