1.10
Technical Safety Checks (TSC)
Abbreviation: TSC = Technical Safety Checks
NOTE
Technical safety checks, test certificate
of the check
1 Identification data
2 General checks
2.1 Visual inspection: Is there any damage
affecting safety?
2.2 Are all cables and cable guides secure
and without any visible damage?
SIREMOBIL Iso-C
•
Within the purview of DIN VDE 0751-1, the operator of medical
engineering products has to perform technical safety checks at
regular intervals.
•
The checks listed in these maintenance instructions contain all
technical safety checks according to DIN VDE 0751-1 except for
the check of the completeness, presence and legibility of the
necessary operator documents.
•
A separate report with print number SP00.000.834.01... is avail-
able for each technical safety check.
•
Within the purview of DIN VDE 0751-1, we recommend docu-
menting the maintenance results both in the maintenance re-
port and in the TSC report. The reports should be filled out
completely and handed over to the client after maintenance is
complete.
•
The table below assigns each technical safety check listed in
document SP00.000.834.01... to the corresponding mainte-
nance work.
SPR2-230.101.02.06.02
07.05
General Information
Maintenance instructions for the
check/comments
Please fill in the fields.
n.a.
3. Mechanical safety, checks:
SIM Covers
SIM Cable deflectors
SIM Foot brake
SIM Brakes
SIM C-arm
Emergency Stop switch
SIM Monitors/displays
3. Electrical safety, check:
SIE Cables and plugs
SIE Voltage discharge rubber
Page 22 of 46