Dini Argeo ALP Manuel opératoire - Page 11

Parcourez en ligne ou téléchargez le pdf Manuel opératoire pour {nom_de_la_catégorie} Dini Argeo ALP. Dini Argeo ALP 19 pages.

4.2 Calibration Menu [.CAL.]

InCal or Span calibration should be performed daily and when the room temperature changes.
InCal [InCal] calibrates the balance using an internal mass.
Span Calibration [Span] uses two weight values: zero and a weight between 50% and 100% of the
capacity of the balance.
Linearity calibration [LIN] uses three weight values zero, 50% of capacity and full capacity. Generally this
calibration is not required unless testing shows that the linearity error exceeds the Linearity tolerance in
the Specification table. (Not available in InCal models)
Calibration Adjust [CAL Adj] allows adjustment to the result of the internal calibration by +/- 99 divisions.
(InCal models only)

4.3 Setup Menu [.setup.]

Automatic Zero-Setting [azsmM] – Environmental changes can cause the display to drift. The Automatic
Zero-Setting Mechanism (AZSM) is designed to keep the balance set at zero, despite these slight
changes. (OFF, SEt.5d, SEt 1d, SEt 2d, SEt 5d)
Filter [filter] – Use the low setting (SEt LO) when environmental disturbances are not present. Use the
medium setting (SEtmMED) for normal environments. Use the high setting (SEt HI) when vibrations or air
currents are present.
Auto Tare [A-tare] – The initial item placed on the balance is assumed to be a container so it is zeroed
out. The next item is then weighed. When the pan is cleared the balance resets, waiting for a container.
(SEt OFF, SEt On)
Grad [Grad] – Select the readability displayed. Reducing the readability may be needed for approval. (SEt
1d, SEt (1)d, SEt 10d)
Global Reset [Gbl rst] – Resets all settings to factory default values. (RESET)

4.4 Mode Menu [.mMode.]

Count Mode [Count] – (SEt On,SEt OFF)
Average Piece Weight (APW) Optimization [aPwW.OPt] – (SEt On, SEt OFF)
Percent Mode [PerCent] – (SEt On, SEt OFF)

4.5 Unit Menu [.Unit.]

The Unit menu is used to enable or disable a specific unit. (SEt ON,SEt OFF) The unit is indicated by a
small character next to Unit in the display (g = grams). The default setting is Grams SET ON and all other
units SET OFF.
T-Units – When unit [t] displays; press Yes to show the T-Unit settings; SEt OFF, Set tt (Taiwan Tael),
SET TH (Hong Kong Tael), SET TS (Singapore Tael), Set tO (Tola) or SEt tI (Tical).
M-Units – When unit [m] is displayed, press Yes to show the M-Unit settings; SET OFF, SEt mMO
(Momme) or SEt mME (Mesghal).
Custom Unit – Custom Unit (C) is used to create a unit of measure not provided with the balance. The
Custom Unit is defined by a factor, a multiplier (E) and a least significant digit (LSD). The balance will use
this to convert grams to a custom unit of measure. (Example: 1 gram = 0.257206 Avoirdupois Dram,
using a 4100g x 0.01g balance)
To create a custom unit, press Yes when unit [c] is displayed.
11