attachMEnt to thE niVcoMP Manual
VeRSion with teSt SeRieS MeMoRy
And Apu-contRoL
tEchnical Data
Storage capacity: 1.800 measurements, dispersed on 1... 8 measurement series
Measurement resolution: 1 mm
interface: wireless (IEEE 802.15.4 / FSK, MSK, range ~25 m)
accEssoriEs
USB-Transceiver-Stick with software FLOOREC
Automatic positioning unit APU-Axx for nivcomp levelling instruments
rEcorD MoDE (suPPlEMEnt to thE niVcoMP instructions)
The version nivcomp RX1 can be used like the standard version of the nivcomp.
In addition the setup of the appliance contains the selection data record.
Data record ON
Data record OFF
rEcorD of lEVElling Data / MoDus Data rEcorD
Start in the recording mode
After [on] the appliance is ready for transmitting and shows the number of the new
record series
storage of the reference level
Displays [ni], then displaying the series number
storage of a measuring point
With keystroke the measuring point (P) of the series is shown
mark a measuring point
During P is displayed, press the button again, the display changes to: +P ...
stoP thE Data rEcorD (lEVElling MoDus)
Press the button about 4 seconds until the display [P #] flashes no longer. This displayed
measuring point will not be stored. The appliance turns itself off automatically.