Assembly
WARNING
To prevent electrical shock or fire, turn off the power switch (flip it to the "
"
side) during assembly.
CAUTION
« + When assembling the microscope, assemble the eyepiece tube first.
+ When moving the microscope, be sure to detach the lamphouse and securely
grasp the arm from the back of the microscope with both hands.
* Do not grasp the coarse and fine focus knobs, eyepiece tube or stage, etc. when
carrying the microscope.
This could cause the part to come off, resulting in
damage to the microscope.
+ Be careful not to pinch your hands or fingers when setting up the microscope.
* Viewing will be adversely affected if any of the lenses is scratched or has fin-
gerprints on it. Handle the lenses carefully during assembly.
* This microscope is a precision optical instrument.
Handle it carefully and do
not subject it to a strong physical shock.
(The accuracy of the objectives in
particular may be adversely affected by even a weak physical shock.)
Assemble each part in sequence as numbered in the diagram.
(For details, refer to pages 25 to 29.)
Tools needed
Two hexagonal screwdrivers (provided)
When not using, place these in the tool holder in the back of the microscope.
Installation location
This product is a precision optical instrument, and using or storing it under unsuitable conditions
may damage it or may have an adverse effect on its accuracy.
The following conditions should
be kept in mind when selecting the installation location.
* Avoid installing the microscope in a brightly lit location such as a room that receives direct
sunlight, or directly under room lights.
The quality of the view through the microscope dete-
riorates if there is excessive ambient light.
¢ Install the microscope in a location that is free from dust or dirt.
* Install the microscope on a flat surface with little vibration.
* Install the microscope on a sturdy desk or table that is able to bear the weight of the instru-
ment.
* Do not install the microscope in a warm, humid location.
In such a location, condensation or
mold may form on the lenses, degrading the performance or damaging the instrument.