In the dialog box, configure the settings as follows.
[Minimum Height]:
Specify the minimum intensity of the bright
points, which was checked in the previous step.
[Maximum Height]: 20000
[CCD Baseline]: 100
[3D]: On
For 2D
[3D]: Off
Click [>>] to make the following detailed settings.
[Minimum Width (nm)]: 200
[Maximum Width (nm)]: 700
For 2D
[Maximum Width (nm)]: 400
[Initial Fit Width (nm)]: 300
[Max Axial Ratio]: 2.5
For 2D
[Max Axial Ratio]: 1.3
[Max Displacement (pix)]: 1
Setting Screening
The bright point images that are excluded by Screening
of Identification Settings are not displayed on any
channel.
Upon the completion of setting, click [OK].
(3)
Click
[Start STORM Analysis].
(4)
Select [Drift Correction], set [Periods] for [1], and
then click [Test].
Test analysis is performed for the currently
displayed period.
Chapter 2
STORM Microscopy Operation
24
5-(2)
5-(4)