Nikon N-STORM Manuel d'utilisation - Page 31
Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Nikon N-STORM. Nikon N-STORM 48 pages. Super resolution microscope
Également pour Nikon N-STORM : Manuel de l'utilisateur (11 pages)
(4)
Click [OK].
The molecules that exceed the threshold are
removed from their channel, and a new channel in
which they are to be stored is created. The channel
is named [destination-channel-name-Xt]. ("Xt"
stands for crosstalk.)
(5)
Check the details of the
[destination-channel-name-Xt] channel tab. If there
is no problem in merging with the destination
channel, right-click the tab and then select [Merge].
To cancel the destination, right-click the
[destination-channel-name-Xt] channel tab and then
select [Undo]. The state existing before crosstalk
subtraction is restored.
(6)
If STORM images are acquired on two channels,
subtract crosstalk for the other channel in the same
way, by following steps (1) to (5).
In this case, perform the steps up to (5) for the first
channel, and then return to (1) and perform the
same processing for the second channel.
7
Click
[File Save] to save the STORM parameters.
The parameters for STORM analysis are saved to the ND2
file.
Type of file used by N-STORM
The file formats used by the N-STORM software are as follows.
File type
ND2 file
Result of STORM
analysis in binary
format
Result of STORM
analysis in text
format
Drift correction data
Open the STORM image on the main screen of NIS-Elements
The currently displayed image can be captured and opened on the main screen of NIS-Elements by clicking
[Create New Elements Document from Current View].
Set ROI and zoom in on it
ROI can be set by clicking
according to the size of the window.
Chapter 2
STORM Microscopy Operation
File name
(ND2-file-name).nd2
(ND2-file-name)_list.bin
(Any-file-name).txt
(ND2-file-name)_drift.txt
[Use ROI]. Right-clicking within the specified ROI range displays the ROI range
6-(5)
Set of conventional images acquired through
NIS-Elements. Parameters for STORM analysis (e.g.,
Identification Settings, Display Options, Filter Settings)
are saved into the ND2 file by selecting the ND2 format
in the [File Save] dialog box of the N-STORM analysis
window.
STORM image (molecule list) created through analysis.
This is automatically created by performing N-STORM
analysis. It can also be saved by using the [File Save]
dialog box in the N-STORM analysis window.
A molecule list can be exported in text format by
selecting text format in the [File Save] dialog box of the
N-STORM analysis window. This is used to share the
results of analysis with other software.
This file is automatically created by performing drift
correction. It is internally used by the N-STORM
software.
27
Description