Cliq AS-1001D Manuel d'utilisation - Page 6

Parcourez en ligne ou téléchargez le pdf Manuel d'utilisation pour {nom_de_la_catégorie} Cliq AS-1001D. Cliq AS-1001D 12 pages. Suction pump

Cliq AS-1001D Manuel d'utilisation

Accessories / Replacement Parts

The following items can be purchased as a replacement for your machine. Please contact our distributors or authorized service centers for inquiries.
Item
Collection container
Intermediate tubing
(Long)
Intermediate tubing
(Short)
Inline filter
Suction tubing
For suctioning of an adult airway, experts recommend sterile, single-use suction catheters with diameters of 10 - 16 Fr/Fg 3.35 - 5.35 mm.
WARNING: Suction catheters are not provided with this device and must be purchased separately, in accordance with the instructions of a trained
healthcare professional or physician:
• Only use disposable, single-use, sterile suction catheters.
• Only use the type and size of suction catheters that are recommended by your healthcare professional.
• Only use suction catheters that are compliant with local and/or national laws, regulations, and directives. In the EU, only use suction catheters that
are labeled with the CE mark, are MDD-compliant, and have received market authorization from a certified notified body. Failure to do so, may cause
harm to the patient.
• Disposable suction catheters are single-use only; do not attempt to re-use or re-sterilize.
WARNING: If third party suction tubing is used, the third party suction tubing must also be compliant with local and national laws, regulations and
directives. In the EU, only use suction tubing that is labeled with the CE mark, is MDD-compliant, and has received market authorization from a
certified notified body. Failure to do so may cause harm to the patient.
Copyright © 2019 ANEST IWATA SPARMAX Co., Ltd. All rights reserved.
Part No.
Specification
3304025C
ø 150 x 185 (H) mm / 800ml
3406073A
7 (ID) x 13 (OD) x 250 (L) mm
3406072A
7 (ID) x 13 (OD) x 150 (L) mm
3406074A
1/4" hose barb x 1/4" hose barb
3318007A
7 (ID) x 10.4 (OD) x 1800 (L) mm
Must be replaced if damaged. One
month expected service life.
Must be replaced if damaged. One
month expected service life.
Must be replaced if damaged. One
month expected service life.
Must be replaced if damaged or has been
in contact with fluid. One month expected
service life.
Must be replaced if damaged. One
month expected service life.
Use only with the same patient. Do not use for multiple patients
6 / 12
Cliq_AS-1001D Manual (230V50Hz_20190125.1)
3510406D