Operation
1.
Operator must wear disposable single-use gloves when operating and handling unit.
2.
Place the machine on a flat and stable surface. WARNING: Inspect power cord. If power cord is frayed or damaged, do not use device.
3.
Prior to use, inspect collection container and tubing for cracks or leaks, and replace all damaged accessories immediately. Do not use device if
accessories are damaged.
4.
Place the collection container securely into the collection container holder.
5.
Connect the inline filter into the inlet, ensuring that the "FLUID SIDE" is facing upwards. Connect one end of the intermediate tubing
to the inline filter on the "FLUID SIDE". Connect the other end of the intermediate tubing to the "vacuum port" on collection container lid.
WARNING: Never operate the unit without the inline filter.
long intermediate tubing
inlet
Disposable suction catheter
(not included with AS-1001D)
6.
Connect one end of suction tubing to the "patient port" on the collection container lid.
7.
Check all connections and make sure they are secure and tight. A poor connection could result in an air leak that diminishes pump performance.
Insert the plug into a wall (mains) outlet with a voltage matching the device's requirements. (See device label located on the underbelly of the
8.
device for device's voltage requirements.) Turn the power switch on.
9.
Adjust the unit to its desired suction level before beginning patient suction.
a. Create a vacuum by completely blocking the open end of the suction tubing.
b. Adjust the vacuum regulator knob until you reach the desired vacuum level displayed on the vacuum gauge.
c. As long as the open end of the suction tubing remains completely blocked, the vacuum level should remain steady.
For suctioning of an adult airway, experts recommend a vacuum level between 100-150 mmHg.
WARNING: Set vacuum pressure in accordance with instructions provided by your healthcare professional. Never exceed your healthcare
professional's maximum pressure guidelines. Failure to follow the instructions and guidance of your healthcare professional could cause bodily
injury.
10.
Once unit is set to the desired vacuum pressure, attach disposable suction catheter to the open end of the suction tubing.
WARNING: Disposable medical suction catheters are not provided with this unit and must be purchased separately in accordance with the
instructions of a trained healthcare professional or physician. Only use suction catheters that are compliant with local and/or national
laws, regulations, and directives. In the EU, only use suction catheters that are labeled with the CE mark, are MDD-compliant, and have received
market authorization from a certified notified body.
11.
Proceed with suctioning as instructed by your healthcare professional. Remove suction catheter from patient immediately after suctioning.
12.
After each use, turn off the power switch and unplug the unit.
CAUTION
• Make sure the power switch is turned off before inserting or removing the plug.
• The unit should not be left running for over 1 hour continuously. If the machine has been operating continuously for 1 hour, please turn off the
power for at least 30 minutes before turning back on.
• Keep the machine and relevant accessories dry. Avoid direct sun exposure during storage.
Copyright © 2019 ANEST IWATA SPARMAX Co., Ltd. All rights reserved.
Connect to suction catheter
7 / 12
inline filter
"FLUID SIDE"
short intermediate tubing
vacuum port
Install inline filter between the collection
container and machine, with the word "FLUID SIDE"
pointed towards the collection container.
Cliq_AS-1001D Manual (230V50Hz_20190125.1)