Baumer PAC50S Panduan Pengoperasian - Halaman 37

Jelajahi secara online atau unduh pdf Panduan Pengoperasian untuk Alat Ukur Baumer PAC50S. Baumer PAC50S 44 halaman.

Baumer PAC50S Panduan Pengoperasian
Baumer
e) Navigate to
f) Enter the given conductivity value of the reference medium in field Conductivity.
g) Click on Calibrate.
Result:
ü
Calibration completed.
INFO
Helpdesk FlexProgram
For more information, see the
Also see about this
2 Reset user calibration [} 39]
11.3.3

Sensor calibration using the Baumer calibration box

In this method, sensor calibration uses calibrated resistors (dry calibration). For doing so, dein-
stall the sensor.
INFO
This task can be done either using the FlexProgram or via IO-Link (e.g. BSS). The following ex-
ample shows calibration with FlexProgram.
Condition:
ð The sensor is connected to a PC with software FlexProgram installed.
ð The sensor is switched on.
ð The calibration box with measurement record is ready for use.
Instruction:
a) Open software FlexProgram on the PC.
b) Open setting PAC50 in the left column.
c) Click on icon
d) Click on button
ü
Dialog window opens.
e) Navigate to
V1 | PAC50H/S
Calibrate
offset.
Help
menu FlexProgram.
Measuring
(Online).
Calibrate Sensor
below the visualization.
Resistance calibration
Operating Manual
section.
Preventive maintenance
|
11