e) Navigate to
f) Enter the given conductivity value of the reference medium in field Conductivity.
g) Click on Calibrate.
Result:
ü
Calibration completed.
INFO
Helpdesk FlexProgram
For more information, see the
Also see about this
2 Reset user calibration [} 39]
11.3.3
Sensor calibration using the Baumer calibration box
In this method, sensor calibration uses calibrated resistors (dry calibration). For doing so, dein-
stall the sensor.
INFO
This task can be done either using the FlexProgram or via IO-Link (e.g. BSS). The following ex-
ample shows calibration with FlexProgram.
Condition:
ð The sensor is connected to a PC with software FlexProgram installed.
ð The sensor is switched on.
ð The calibration box with measurement record is ready for use.
Instruction:
a) Open software FlexProgram on the PC.
b) Open setting PAC50 in the left column.
c) Click on icon
d) Click on button
ü
Dialog window opens.
e) Navigate to
V1 | PAC50H/S
Calibrate
offset.
Help
menu FlexProgram.
Measuring
(Online).
Calibrate Sensor
below the visualization.
Resistance calibration
Operating Manual
section.
Preventive maintenance
|