BD FACSCanto II Panduan Referensi Cepat
Jelajahi secara online atau unduh pdf Panduan Referensi Cepat untuk Alat Ukur BD FACSCanto II. BD FACSCanto II 7 halaman. Flow cytometer
BD FACSCanto II Flow Cytometer
This guide contains instructions for both BD FACSCanto™ clinical and
BD FACSDiva™ software. Follow only those sections that apply to the
software you are running.
Starting Up
1
Turn on the cytometer main power.
2
Start up the computer, launch the software, and
log in.
3
Make sure the software is connected to the cytometer; if needed,
connect manually (choose Cytometer or Instrument > Connect).
4
Check fluid levels in the Status or Instrument window.
Empty the waste or replace fluid containers, if needed.
5
If automatic cleaning is disabled, run the fluidics startup procedure.
To prevent fluid overflow, make sure there is no tube on the sample
injection tube (SIT) during fluidics startup.
6
Check the flow cell for air bubbles; if needed, run the de-gas flow cell
procedure.
7
Check that laser warmup has finished.
Quick Reference Guide
BD FACSDiva software
BD FACSDiva software
1
or
BD FACSCanto clinical software
BD FACSCanto clinical software