Shutting Down: BD FACSCanto Clinical Software

1
Create a worklist with a two-tube panel; enter Clean as the Sample ID.
2
Click
; when asked to save the worklist, click No.
3
(Loader option only) Click
When prompted, install a tube containing ≤3 mL of BD™ FACSClean
4
solution or 10% bleach.
5
Click
6
When prompted, unload the tube.
Install a tube with ≤3 mL of DI water; repeat steps 5 and 6.
7
To prevent fluid overflow, do not leave a tube on the SIT during fluidics
shutdown.
8
Choose File > Exit, and select Run fluidics shutdown and exit.
9
When fluidics shutdown completes, turn off the power.

Shutting Down: BD FACSDiva Software

Install a tube with ≤3 mL of BD FACSClean solution or 10% bleach,
1
and click
2
After 5 minutes, click
3
Repeat steps 1 and 2 with DI water.
4
Run fluidics shutdown.
To prevent fluid overflow, do not leave a tube on the SIT during fluidics
shutdown.
5
Turn off the instrument main power.
. Wait 5 minutes, and then click
.
5
to run tubes manually.
.
and remove the tube.