BD Alaris GW 800 Series Panduan Referensi Cepat - Halaman 11
Jelajahi secara online atau unduh pdf Panduan Referensi Cepat untuk Pompa Air BD Alaris GW 800 Series. BD Alaris GW 800 Series 12 halaman. Volumetric pump

Advisories alert the user and are indicated by an audible alarm, a message on the
display or both.
Check the display for advisory message.
Press
to silence the sound.
c
Advisory / Condition
C
Bolus is being administered.
D
VTBI limit reached.
E
The pump is priming the infusion set.
G
The pump is on hold.
I
Automatic infusion set check.
Action
• Release
b
infusion once prescribed bolus has
been administered.
• Pump will infuse at the KVO rate,
until
is pressed.
d
• Ensure all air has been primed out of
the infusion set, before starting the
infusion.
• Press
to restart / continue infusion
d
or press
e
• Allow test to complete before
operating the pump further.
0000CF04392 Iss. 1 11/12
key to continue the
to return to setup.