Advisories alert the user and are indicated by an audible alarm, a message on the
display or both.
Check the display for advisory message.
Press
to silence the sound.
c
Advisory / Condition
C
Bolus is being administered.
D
VTBI limit reached.
E
The pump is priming the infusion set.
G
The pump is on hold.
I
Automatic infusion set check.
Action
• Release
b
infusion once prescribed bolus has
been administered.
• Pump will infuse at the KVO rate,
until
is pressed.
d
• Ensure all air has been primed out of
the infusion set, before starting the
infusion.
• Press
to restart / continue infusion
d
or press
e
• Allow test to complete before
operating the pump further.