CONTROL SOLUTION TEST

Material needed:
● Glucose Meter
● Test strips
● TaiDoc control solution
When to do a control solution test:
◆ When you suspect that the meter or test strips are not
working properly.
◆ If your blood glucose test results are not consistent with
how you feel.
◆ After the test strips are exposed to extreme conditions.
◆ After dropping the meter.
The control solution is to check that the meter and the test
strips can work together as a system and work correctly.
Step 1
Insert the test strip to turn on the meter. Insert the test strip into
the meter. Wait for the meter to display the drop symbol.
Step 2
You may press "S " button to select CTL mode, for
selecting the CTL mode, please refer to the "THE
TWO MEASURING MODES" section.
Step 3
Apply the control solution
Shake the control solution vial thoroughly before use.
Squeeze out a drop and wipe it off, then squeeze another drop
and place it on the tip of the vial cap.
Hold the meter to move the absorbent hole of test strip so it
touches the drop. Once the confirmation window fills completely,
the meter will begin counting down.
To avoid contaminating the control solution, do not directly
apply control solution onto a strip.
The result of the control solutions test appears on the display.
The result will not be stored under CTL mode.
The result of the control solutions test should be within the range
printed on the test strip vial label. If the test results falls outside
the specified range printed on test strip vial label, repeat the test.
Results falling outside the specified range may be caused by:
■ Error in performing the test
■ Expired or contaminated Control Solution
■ Expired or contaminated Test Strip
■ Malfunction of your Meter
If you continue to get control solution test results that fall outside
of the range printed on the vial, the system may not be working
properly, Do not use the meter. Please contact your distributor
support immediately
NOTE:Do not use the expired control solution.Only use the
TaiDoc control solution, you can purchase this product
through visits the website: www.taidoc.com