13. Continue testing or switch off the
instrument as required. Unplug the power
supply if the instrument is not to be used for
long periods.
Operation Notes:
The instrument will accommodate panels
with a maximum width of 100mm.
Reference to the test method recommends
the use of burnished steel panels with a
width and length of at least 70 mm and
a thickness range of 0.3mm to 1.25mm
(maximum).
Do not exceed the above fi gures as this may
cause permanent damage to the instrument.
Zero
All measurements are taken from the
zero position (where the indenter is level
with the base of the clamp area). It is not
necessary to re-zero the zero position when
power is switched off and on. However,
large ambient temperature changes will
affect the readings to a small extent, and
it is good practice to regularly re-zero the
instrument.
Power Off
Turn off the power using the switch on the
back panel. The zero position will be stored
during power off.
Operation
13