7.4. Rotor Nut
If the rotor Nut needs to be tightened, use a ½" nut driver and tighten it by hand
until snug.
Rule of thumb: One full thread showing above the nut is a good indicator of proper
installation.
7.5. Speed Calibration
Check the centrifuge speed periodically, every two years is recommended.
Important: When verifying rotor speed, make certain that all carriers are installed
in the rotor.
No calibration adjustment of speed can be made, only a verification of rotor speed.
7.6. Removing the Cabinet (Upper Housing)
There are nine screws that fasten the centrifuge cabinet to the base.
Begin by unplugging the centrifuge and waiting 10 minutes for internal voltages to
dissipate.
Use a #2 Phillips screwdriver to remove the cabinet screws (six on the bottom,
three in the rear)
The centrifuge control panel is attached to the base internally with cable
harnesses. Be careful not to stress the cables when removing the cabinet.
Lift the cabinet straight up and off the base, setting it down on its front side.
Gently remove the combination power supply / motor harness from the control
board.
7.7. Replacing the Lid Lock Assembly
The lid lock assembly is accessible once the cabinet has been removed.
Gently remove the lid lock wire harness' from the control board.
The lid lock assembly is held in place with four screws.
Two #8-16 screws on the inside and two #8-32 screws on the outside (covered by
the open/close label).
Remove the Open/Close label to access the two screws underneath.
To install, line up the holes on the lid tray assembly to the screw holes on the
cabinet and install the four screws.
Complete the installation by plugging the wire harness into the control board
header 'J2'.
DRUCKER DIAGNOSTICS
HORIZON 6 SERVICE MANUAL
SM024
REV A
6