4. Optional: Analogue In- / Outputs
Devices of the DF-C series can optionally be equipped with analogue inputs / outputs for setting and reading the
output parameters. The analogue outputs / inputs are accessible via a labeled terminal block which is located either
on the back, the front or behind the service door (depending on device model). The pin assignment is explained on the
terminal block for a simple and straight forward connection.
4.1 Optional: Ext. Output On/Off Ctrl. & Interlock
Input for controlling the status of the output of the laboratory power supply (on / off), switchable as "Interlock" or
external control. This input is configured as a two pin connection, a „true" state is triggered by shorting the two pins
and a „false" state is triggered by removing any connection between the two pins of the input.
This option can be configured as either an interlock input, which disables the output of the power supply if the state is
false, or as an external output status control which enables or disables the output of the power supply depending on
the control signal state (true = on/false = off) if the „Output On/Off" switch of the power supply is always in the ON
position.
4.2 Optional: Potential-free output / configurable output
The output of DF-C series devices in the basic version has a potential to PE, which follows the usual protection
regulations. All models can also be ordered in the following modifications (please see the „DF-C Series Function and
Wiring Diagram" PDF for details):
1). Configurable output without ELCB
The output of the power supply has no potential to the input (PE), so the output of the device has no protective earth
for loads that may require a PE connection and the load is not protected by a ELCB ! By bridging output N to input N,
the output can be pulled to the input potential, in this case input PE = output PE and the load is protected by the ELCB
installed in the operation environment (Laboratory).
2). Configurable output with built-in ELCB
The model with integrated ELCB offers the following switchable configurations:
A). Potential-free output with its own ELCB
Connections output
L (1, 2, 3): Phase (potential-free)
N: Neutral conductor (potential-free)
GND: In this configuration to use as PE connection for the load
Remark: ELCB triggers as soon as a residual current flows through the GND conductor.
Connections input
L1: Input phase 1
L2: Input phase 2
L3: Input phase 3
N: Neutral conductor
PE: Earth conductor
DSC-Electronics Germany • Georgstraße 36 • 53111 Bonn