Cleaning
Troubleshooting
Symptom / Display
Cannot turn on
Weight value blinking
"Tare" shown
"Close Cover" shown
Poor accuracy
Cannot calibrate
Err 7.0
Err 8.1
Err 8.2
Err 8.3
Err 8.4
Err 9.5
Err 53
Service Information
If the troubleshooting section does not resolve or describe your problem, contact your authorized DSC service agent. Please visit our
web site: www.dsctest.com. Call 800-726-5883
Accessories
Description
Sample Test Accessories (100 Test)
Sample Pan Handler
50g Calibration Weight
Thermal Printer
Paper for Printer
Disconnect the Fat tester from the power supply before cleaning the Fat tester.
Make sure that no liquid enters the interior of the Fat tester.
Make sure the Fat tester is cooled down before cleaning.
Possible Cause
No power to Fat tester
Sample weight less than 0.5g
Pan weight needs set to zero
Cover needs to be closed before starting test
Improper calibration
Unstable environment
Unstable environment
Incorrect calibration masses
Time out
Pan Support has load during power on
Pan Support was removed prior to power on
Weight on Pan Support exceeds capacity
Pan Support was removed during weighing
Factory calibration data corrupted
EEPROM Checksum error
Part No.
PD100T
80252476
51054-16
DSC103G
PP250X80
Clean the Fat tester at regular intervals.
IR Coil
Housing surfaces and the temperature sensor
Temperature sensor
may be cleaned with a lint-free cloth slightly
dampened with water or a mild cleaning agent.
Glass surfaces may be cleaned with a
commercial glass cleaner.
Do not use solvents, harsh chemicals, ammonia
or abrasive cleaning agents.
Remedy
Verify connections and voltage
Increase samle size
Press Tare
Close Cover
Perform calibration
Move the Fat tester to suitable location
Move the Fat tester to suitable location
Use correct calibration masses
Remove weight from pan support
Install Pan Support
Remove weight from the Pan Support
Re-install Pan Support
Contact DSC at 800-726-5883
Contact DSC at 800-726-5883
Description
Printer Ribbon Cassette
Protective Shipping Casing
Carrying Case
Security Locking Cable
Security Lock (Kensington® type)
Temperature Calibration Kit
8
Part No.
RC200P
200S
TBD
76288-01
470004-010
11113857