QS11 Operator Notes
Sectioning – The recommended blade holder angle for most applications is 10°
After each use – Remove any debris from the microtome and cryo-chamber. Especially
clean the blade holder, specimen head and vertical travel areas. A build-up of debris
over time may freeze causing "knocking" or "thumping" during handwheel operation.
Periodically – Empty the liquid that accumulates in the plastic container from the defrost
cycles. The container should be periodically disinfected.
Defrosting – Due to the extremely low temperature in the cryo-chamber and the window
remaining open for extended periods of time during operation, humidity from the room
will create a frost build-up inside the cryostat. It may be necessary to occasionally turn
the cryostat off and perform a complete defrost of the unit.
QS11 Troubleshooting Guide
Issue
Error 1
Error 2
Error 3
Error 4
"H" Codes
Keypad unresponsive
* Warning: The cabinet and microtome must be entirely dry before cooling down the cryostat.
Any moisture inside or outside the microtome will freeze. This may result in a malfunction or
possible damage to the instrument.
Remedy
Electronics issue – contact technical assistance
Software backup battery needs to be replaced
Cabinet temperature is above 0°C. Confirm the microtome is dry*. Press the
"Memo" button to proceed
Confirm the cable to the microtome is properly connected
This is not an error but an indication that the cryostat is in a defrost cycle
A power failure occurred. Turn the main power switch off, wait 30 seconds and
restart the cryostat. Confirm the cabinet temperature is below 0°C and that the
microtome is dry*.
Avantik Technical Assistance: 1-800-783-9424