Abaxis VetScan HM5 Panduan Referensi Cepat - Halaman 10

Jelajahi secara online atau unduh pdf Panduan Referensi Cepat untuk Alat Ukur Abaxis VetScan HM5. Abaxis VetScan HM5 19 halaman. Hematology analyzer
Juga untuk Abaxis VetScan HM5: Panduan Referensi Cepat (17 halaman), Panduan Instalasi (1 halaman), Petunjuk Instalasi (2 halaman)

Abaxis VetScan HM5 Panduan Referensi Cepat
Maintenance
(14 days)
Soak Cleaning should be performed every 14 days. When the software is upgraded to v2.3
or higher, the Soak Clean Interval will automatically be defaulted to 14 days.
If Scheduled Maintenance is Enabled, a message will appear when a cleaning is due. Touch
Next, then follow the on-screen instructions. Alerts will start appearing 3 days before the
cleaning is due. Maintenance is required on the due date.
Instructions will guide the user through the cleaning. This cleaning takes 10 minutes; after the
cleaning is completed run a Blank, and if the Blank is acceptable, proceed with Quality control.
If your HM5 analyzer is not set up for Soak Cleanings every 14 days, refer to HM5
Operator's manual.