Maintenance
(14 days)
Soak Cleaning should be performed every 14 days. When the software is upgraded to v2.3
or higher, the Soak Clean Interval will automatically be defaulted to 14 days.
If Scheduled Maintenance is Enabled, a message will appear when a cleaning is due. Touch
Next, then follow the on-screen instructions. Alerts will start appearing 3 days before the
cleaning is due. Maintenance is required on the due date.
Instructions will guide the user through the cleaning. This cleaning takes 10 minutes; after the
cleaning is completed run a Blank, and if the Blank is acceptable, proceed with Quality control.
If your HM5 analyzer is not set up for Soak Cleanings every 14 days, refer to HM5
Operator's manual.