Olympus BX52-P Petunjuk Manual - Halaman 26
Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Mikroskop Olympus BX52-P. Olympus BX52-P 36 halaman. Polarizing microscope
Juga untuk Olympus BX52-P: Petunjuk Manual (34 halaman)
Fig. 31
23
3
Adjusting the Eyepiece Cross Lines
}To align the eyepiece cross lines and the vibration direction, the
orientation plate provided with the U-CPA or U-OPA is required.
Orientation plate (U-PJ: Single module name)
# Remove the test plate adapter and the compensator from the light
path.
1. Swing out the top lens of the condenser.
2. Place the orientation plate on the stage. Remove the analyzer from the
light path, and use the 4X objective to focus.
(Fig. 31)