Olympus BX52-P Manual de instruções - Página 26
Procurar online ou descarregar pdf Manual de instruções para Microscópio Olympus BX52-P. Olympus BX52-P 36 páginas. Polarizing microscope
Também para Olympus BX52-P: Manual de instruções (34 páginas)
Fig. 31
23
3
Adjusting the Eyepiece Cross Lines
}To align the eyepiece cross lines and the vibration direction, the
orientation plate provided with the U-CPA or U-OPA is required.
Orientation plate (U-PJ: Single module name)
# Remove the test plate adapter and the compensator from the light
path.
1. Swing out the top lens of the condenser.
2. Place the orientation plate on the stage. Remove the analyzer from the
light path, and use the 4X objective to focus.
(Fig. 31)