5

USING THE CONTROLS

5-1 Base

1
Fig. 11
Fig. 12
1
Adjusting the Brightness
Turn the brightness adjustment knob @ clockwise to increase the voltage
and make illumination brighter.
} When the brightness adjustment knob @ is adjusted to the camera
symbol marking, the brightness of the light becomes the daylight (color
temperature of about 5500K). However, note that the daylight may not
always be reproduced accurately depending on the degree of bulb
consumption, etc.
2
Using the Light Intensity Preset Switch
The light intensity preset switch @ makes it possible to limit the light
1
intensity to a preselected level regardless of the position of the brightness
adjustment knob.
2
The light intensity preset has been preset to optimum brightness for
photography when the built-in LBD filter is engaged (approx. 9 V with
marking).
1. Press the light intensity preset switch @ to the ON position. (The face of
the switch lights when it is ON.)
2. Using a small flat-blade screwdriver, turn the preset adjustment screw 2 to
obtain the required light intensity. Turning the screw clockwise increases
brightness.
3. When the light intensity preset switch @ is set to OFF, the brightness
returns to the level set by the brightness adjustment knob.
} While the light intensity preset switch is ON, turning the light brightness
adjustment knob does not affect brightness.
BX53
(Fig. 11)
(Fig. 12)
13