Olympus BX60 Petunjuk Manual - Halaman 28

Jelajahi secara online atau unduh pdf Petunjuk Manual untuk Mikroskop Olympus BX60. Olympus BX60 48 halaman. System microscope
Juga untuk Olympus BX60: Panduan Memulai Cepat (4 halaman)

Olympus BX60 Petunjuk Manual
BX60
Rotating the Stags
(Rg.43)
1. Slightly loosen the stage damping screw ©.
2. The stage can be rotated both to the left and to the right by turning
it with the stage clamping screw.
* A dick may be heard and fett during rotation. However, this is
duo to the construction of the stage bfsckot and does not indicate
a malfunction.
O The rotation angle changes depending on the position of the stage
knobs.
O The U-SIC stage cannot be rotated.
Stage Height Adjustment
(Rgs.44,46)
O By lowering the position of the substage. the microscope will
accommodate specimens with maximuin heights of 40 mm. This is
useful when observing metallurgical specimens end other thick objects.
1. Lower the stage to its tower Emit then remove the stage from the
microscope. (See page 3)
2. Using the Allan screwdriver, loosen the substage bracket clamping
screw © and remove the substage. (Fig. 44)
3. Turn the coarse adjustment knob and raise the focusing block (J) to
where the stopper screw (D in the ami becomes visible. (Fig. 4S1
4. Using the Allen screwdriver, loosen and remove the upper stopper
screw®.
5. Reattach substage bracket and stage.
O Store the removed stopper screw ® in a safe place so that it win not
be lost il needed again.
24